Product Details
OW-8000A multi-parameter patient monitor is our newest product. The display is 7 inch high resolution color TFT. It is equipped with 4 parameters including ECG, NIBP, SpO2, Resp. These four parameters can be flexible. You can choose different combinations as you need. This monitor is built-in rechargeable lithium battery. The working capacity is 2 hours.
Features
1. 4 Standard parameters: ECG, NIBP, SpO2, Resp
2. 7 inch high resolution color TFT LCD display
3. Real time S-T segment analysis, pace-maker detection
4. Indication of incorrect manipulation and malfunction analysis
5. Patient info input management function
6. Multi-lead ECG waveforms display in phase
7. Large volume of tabular and graphic trends information storage and easy to recall
8. Capture dynamic waveforms
9. Efficient resistance to interference of defibrillator and HF knife
10. Up to 2 hours working capacity of built-in rechargeable lithium battery
11. Unique accessory management cabinet
12. Networking capacity
13. Option: Temp, EtCO2, Built-in recorder, Cart, Wall shelf.
Specification |
ECG Specification |
Standard 3-lead or 5-lead |
3-lead RA,LA,LL,Lead method:I,II,III 5-lead RA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V |
Gain |
×250, ×500, ×1000, ×2000 |
HR Measuring Range |
Adult:15 ~ 300bpm (beat/minute) Neonatal baby/children:15 ~ 350 bpm(beat/minute) |
HR Accuracy |
±1% or ±1bpm,the larger prevails |
HR Resolution |
1 bpm(beat/minute) |
Sensitivity |
> 200 uV P-P |
Differential Input Impedance |
> 5M ohm |
Bandwidth |
Diagnostic Mode: 0.05~130 Hz Monitoring Mode: 0.5~40 Hz Operation Mode: 1~20 Hz |
CMRR |
Diagnostic Mode: > 90 dB Monitoring Mode: >100 dB Operation Mode: >100 dB |
Electrode Offset Potential |
±300 mV |
Leakage Current |
<10 uA |
Pacing Pulse Test |
Test pacing pulse in accordance with the following conditions: Range: ±2~±700 mV Width: 0.1~2 ms Rise time: 10~100 us |
Pacing pulse inhibition |
When pacing analysis switch is on, pacing pulse in accordance with the following conditions are restrained, but affection against HR calculation. Range: ±2~±700 mV Width: 0.1~2 ms Rise time: 10~100 us |
Baseline Recovery |
< 3s After defibrillation |
Signal Range |
±8mV P-P |
Calibrating Signal |
1mV(Peak-to-peak value),precision:±5% |
ST Segment Measuring Volume |
Measuring Range: |
-2.0mV ~ +2.0mV |
Measuring precision: |
Ranging -0.8 mV ~ +0.8mV, measuring error is ±0.02mV or ±10%,The larger prevails. No definition for other ranges. |
Resp Specification |
Measuring Method |
RA-LL impedance |
RESP Impedance Measuring Range |
0.3~3Ω |
Base Impedance Range |
200~4000Ω |
Bandwidth |
0.1~2.5Hz |
RR Measurement Range |
Adult: 0~120BrPM Children and neonatal baby: 0~150 BrPM |
RR Resolution |
1 BrPM |
RR Precision |
±2 BrPM |
Asphyxia Alarm |
10~40 seconds |
NIBP Specification |
Measuring Method |
Pulse wave oscillometry |
Work Mode |
Manual/Automatic/STAT |
Measuring Interval of Automatic Measuring Mode |
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s) |
Measuring Time of STAT Mode |
5 minutes |
PR range |
40 - 240 bpm |
Measuring Range |
Adult Sys: 40~270mmHg Dia: 10~215mmHg Mean: 20~235mmHg Children Sys: 40~200mmHg Dia: 10~150mmHg Mean:20~165mmHg Neonatal baby Sys: 40~135mmHg Dia: 10~100mmHg Mean: 20~110mmHg |
Static Pressure Range |
0~300mmHg |
Static Pressure Precision |
±3mmHg |
Pressure precision |
Max. average error: ±5mmHg; Max. standard deviation: 8mmHg |
Over voltage protection |
Adult mode: 300 mmHg Children mode: 240 mmHg Neonatal baby mode: 150 mmHg |
SpO2 Specification |
SpO2 Measuring Range |
0~100% |
SpO2 Resolution |
1% |
SpO2 Precision |
70~100%: ±2 DIGIT 0%~69%: no definition given |
PR Measuring Range |
20~300bpm |
PR Resolution |
1bpm |
PR Precision |
±3bpm |
Temp Specification |
Temperature Sensor |
YSI series,CYF series |
Channel Quantity |
2 Channels |
Measuring Range |
0~50ºC |
Resolution |
0.1ºC |
Precision |
±0.1ºC(Excluding Sensor Error) |
IBP Specification |
Label |
ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
Measuring and alarm range |
ART: 0~300 mmHg PA: -6~120 mmHg CVP/RAP/LAP/ICP: -10~40 mmHg P1/P2: -10~300 mmHg |
Press Sensor Sensitivity |
5 uV/V/mmHg |
Press Sensor Impedance |
300-3000Ω |
Resolution |
1 mmHg |
Accuracy |
±2% or±1mmHg, which great |
Actualization interval |
about 1 Sec. |